User`s guide

Installation and Setup
79
may need to make minor adjustments. Use this interactive process
until you obtain the desired flow rates.
An Inlet Split flow of at least 25 mL/min should be established even
if inlet split is disabled during analysis. After desorption and cool-
down, and before being unloaded, sample tubes are depressurized
using the inlet split function.
The instructions below are for both the TD and ATD. You must load
a tube into position on the carousel of the ATD that will
subsequently be loaded into the instrument to complete the flow
path. If you are using the TD, you must load the tube into the
instrument directly.
1. Press the power switch located on the rear panel to
switch on the instrument.
2. To prepare the TD for setting up the Desorb flow rate
and Inlet Split flow rate, a sample tube must be loaded
and sealed into the desorb position. If you are using the
TD, load a tube. If you are using the ATD, ensure a tube
is available in position #1 on the carousel.
NOTE: Use a clean sample tube that does not contain a sample for
the set-up procedure. The sample tube should contain the
type of packing material that will typically be used for your
applications.
3. When the ATD/TD displays a Ready status, open the
Tool s menu and select Adjust Flows.
4. Select Inlet Split.
5. The ATD will load the tube from position #1 into the
sampling position to complete a carrier gas path. Once
the tube is loaded, the heated valve is rotated, and the
solenoid valves, which control the flows are opened.
6. To set the inlet split flow, connect the flow meter to the
split vent. Adjust the inlet split needle valve to obtain
the required flow.
CAUTION
Carrier Gas and Dry Air must be already connected to the
TD and flowing for a few minutes.