User`s guide
Installation and Setup
77
the instruments cold trap. To install the dryer assembly use the
instructions provided with this assembly (Part No. N6710048).
Setting the Carrier Gas Flows
When you set the carrier flow you are setting the flow for the
desorption system which includes purging the sample tube, primary
and secondary desorption, inlet and outlet split flows and the GC
column flow rate.
Relatively high flow rates (20 to 100 mL/min) are required for
purging and desorption. Substantially lower flow rates (1 to 5 mL/
min) are required for capillary column chromatography. The
multiple splitter allows you to set the required desorption rates and
provides the lower column flow rates required by the GC. If your
application requires splitless operation you will need to install a
wide bore capillary column or a packed column in your GC in order
to handle the higher flows required for desorption.
You must set the Inlet Split flow manually using the Inlet Split
needle valve on the ATD/TD. You will set the Outlet Split flow
separately using the Outlet Split needle valve.
You must set the carrier pressure first. Then you will set the desorb
and inlet split flows. Lastly, you will adjust the outlet split flow rate.
You must power up the instrument before you can set the carrier
pressure and the various flow rates. Refer to Chapter 3 for
instructions on powering up and operating the ATD/TD.
Setting the GC Column Head Pressure
CAUTION
Do not power up the GC until you have established carrier
gas pressure and very dry air for at least 20 minutes. Dry Air
supply must be connected to the TD and air is flowing. The
cold trap and Peltier should be in an absolutely dry
environment before power is applied. Especially if a new unit
is just delivered or was off for a long period of time in a
humid area.










