Column Leak Test
1. To perform the column leak test, cap one end of the
column and/or transfer line with a rubber septa. Connect
carrier gas, and pressurize the system.
2. Go to the Tools drop down menu and select
Maintenance and select Column Leak Test. The
following screen will appear. Press OK to start the test.
Check for leaks or pressure drop
3. If your GC fails this column leak check all connections
and ferrules using a leak detector or a 50/50% IPA/water
mixture with a swab or pipette.
4. If your GC passes this check you have successfully
connected the capillary column to the TD.
Connecting the Dryer Assembly
The dryer assembly will reduce the amount of moisture entering the
instrument. It is highly recommended that you use the dryer
assembly (Part No. N6710048) to remove moisture that may damage