User`s guide

Installation and Setup
72
Use only carrier gases with a purity of 99.995% or better. Only top
quality gases are suitable; typical laboratory supplies are usually not
pure enough for thermal desorption. Gas purity can be improved
when filters are included in supply lines. Hydrocarbon, oxygen and
moisture filters are recommended for carrier gas lines.
Figure 13 Connect Filters to the Gas Supply System
When oxygen filters and moisture filters are used together, install
them in the gas line in the following order: gas cylinder, hydrocarbon
filter, moisture filter, oxygen filter, and TD. This sequence prevents
any hydrocarbons present in the gas stream from reaching the
oxygen filter.
Most filters are disposable. Replace them when a new cylinder is
installed or as soon as contamination is suspected.
The TurboMatrix 300 TD, TurboMatrix 350 ATD and the
TurboMatrix 650 ATD instruments have separate gas connections
for tube and trap desorb. This allows you to use different gases (He,
H
2
or N
2
). If you wish to use the same gas for both the trap and
desorb sue the connection N671-0105 provided with the shipping
kit.
To connect carrier gas to your system:
NOTE: Ensure that carrier gas is connected to the instrument
before powering on the instrument.
1. Locate all carrier and dry air tanks in a secure location.
WARNIN
G
Ensure the pressure rating of the filters has not been
exceeded.
TD
Moisture
Filter
Oxygen
Filter
Charcoal
Filter
Carrier
Gas
Cylinder