User`s guide

Installation and Setup
68
0497-2066) or the graphite ferrule supplied with the
union (not for ECD or MS). After passing the column
end through the ferrule cut off a further 0.5cm of fused
silica to avoid the possibility of a blockage. The silica
line should end in the center of the union. It is essential
that the fused silica line does not touch any part of the
oven wall.
NOTE: The ferrules are fitted with the tapered end away from the
body of the union. Tighten the nut just enough to avoid
leakage.
8. Connect the packed column to the injector using two
wrenches to avoid stressing the braze assembly.
Direct Connection of a Capillary
Column to the TD
In this arrangement the front end of the GC column is passed though
the heated transfer line and connected directly to TD. If you
configure your instrument in this manner, you must ensure that the
column coating material is able to withstand the temperatures of the
heated transfer line and the heated valve.
To install your capillary column for direct connection:
1. Install the insulated transfer line through the rear panel
or through the top of the GC as outlined in Installation
Through the GC Rear Panel on page 53 or Installation
Through the GC Injector on page 55.
2. Mount the column on the capillary support assembly.
NOTE: It is essential that the column does not touch any part of the
oven wall.
3. Unwind a sufficient length of the column to pass
through the stainless steel tube (approximately 1.5 m).
Thread the end of the capillary column into the exposed
stainless steel tube at the GC end so that it protrudes
from the transfer line at the ATD/TD end.
4. Install the capillary column at the ATD/TD as outlined
in Installation at the ATD/TD on page 57.