User`s guide

Installation and Setup
55
Figure 6 Location of the Access Hole on the AutoSystem XL and
Clarus 500 GC Rear Panel
Installation Through the GC Injector
To install the heated transfer line, through the top of the GC:
1. Disconnect the GC from any electrical or gas supply.
2. Two right-angle bends with a radius of approximately 8
cm (3.25”) should be put in the transfer line assembly.
At the ATD/TD end, the bend should leave 9 cm of
straight tube at the end and the electrical connections
should be pointing downward. See Figure 7.
3. At the GC end, a bend leaving a straight length of 50 cm
in the same plane and direction should be made.
WARNIN
G
Do not make the bends too tight as you will damage the
heater. It will also be difficult to feed the fused silica line
through the stainless steel sleeve.
Transfer Line
Rear Panel of the AutoSystem XL
Carr
Carr
2
Air
2
Air
M/U
or H2
M/U
or H2
Transfer Line
Access Hole
or Clarus 500