User`s guide

Installation and Setup
53
Normally you will install the outer components of the transfer line at
the GC and then at the TD. Lastly, you will run either the fused silica
transfer line or the analytical column, through the transfer line, from
the GC oven to the ATD/TD and make the required connections.
Installation Through the GC Rear Panel
Clarus 500 or AutoSystem XL—A cut out for the ATD/TD transfer
line is located on the rear panel of the AutoSystem XL. If the GC
PPC module is installed, then the rear panel has been replaced with
the PPC option and you must install the transfer line through an
empty injector or detector port on top of the instrument
Other GCs—If you are using another type of GC, you can use
another access hole on the rear panel. Alternatively, you can use an
empty injector or detector port on top of the instrument. See
Installing the Transfer Line Through the Top of the GC on page 64.
To install the heated transfer line, through the back of the GC:
1. Disconnect the GC from any electrical or gas supply.
2. You must bend the transfer line to fit between the rear of
the GC and the top of the ATD/TD. Observe a minimum
radius of 6 cm.
3. At the GC end, a bend leaving a straight length of
approximately 50 cm in the same plane and direction
should be made.
4. Position the ATD/TD to the right side of the GC. The
distance between the two instruments should be about
15 cm.
5. Feed the transfer line through the rear panel access hole
and then carefully remove the rear panel of the GC. The
1-inch access hole in the top right-hand corner of the
WARNIN
G
Do not make the bends too tight, as you will damage the
heater. It will also be difficult to feed the fused silica line
through the stainless steel sleeve.