User`s guide
Installation and Setup
39
Electrical Connections
AC Line Connections
Table 7 Power Requirements
Additional equipment, such as computers and printers, should be
connected as per their specifications.
Instruments and peripherals should not be connected to or near
circuits with large inductive or large variable loads (i.e. large motors,
discharge lamps, photocopy systems, radio transmitters, etc.). All
instruments of the system should be connected to a common phase.
The line power supply should conform to local safety regulations
and must include a correctly wired protective earth/ground terminal.
Power
Consumption
Approximately 1000 VA for the ATD/TD. The
AutoSystem XL and Clarus 500 GC will consume
2400 VA
Power
Specifications
All electrical supplies must be smooth and free of
transients greater than 40 volts peak-to-peak and
must meet and remain within the following
tolerances:
ATD/TD 100V±10%, 8A @ 50/60 Hz ±1%
120V±10%, 8A @ 50/60 Hz ±1%
220V±10%, 4A @ 50/60 Hz ±1%
230-240V±10%, 4A, @ 50/60 Hz
±1%
AutoSystem
or
Clarus 500
120 VAC ±10% @ 50/60 Hz ±1%
230 VAC ±10% @ 50/60 Hz ±1%
Power
Outlets:
ATD/TD A minimum of one dedicated 120
VAC outlet at 15 amps or a 230 V
outlet at 6 amps
AutoSystem
or
Clarus 500
A minimum requirement of one
dedicated 120 VAC outlet at 20 A or
one 230 VAC outlet at 10 A or
greater.