User`s guide

Installation and Setup
36
Table 5 Required Laboratory Conditions
Altitude
(operating)
Sea level (0 metres) to 2000 m (6,562 feet)
Storage
(non-
operating)
You can store the instrument safely under the
following conditions:
Ambient temperature –20 to 60 ºC (-4 to
140 °F)
Ambient relative humidity of 20% to 75%,
(non-condensing)
Altitude in the range 0 m to 12000 m
(39,370 feet)
When you remove the instrument from storage,
allow it to stand for at least a day under the
approved operating conditions before putting it
into operation.
Temperature
(Safe
Operating)
+ 5 ºC to 40 ºC (41 ºF and 104 ºF)
Installation This instrument is able to withstand transient
over-voltage according to Installation Category
II as defined in IEC 61010-1 and IEC 60664
Other The laboratory should be free of flammable,
explosive, toxic or corrosive vapors.
Always provide adequate ventilation. When
analyzing hazardous compounds, such as
pesticides, it may be necessary to arrange for
venting the detector effluent into a fumehood.
Gas cylinders should be located outside of the
laboratory whenever possible. All gas cylinders
should always be stored and operated in the
vertical position and should be firmly clamped
to a suitable surface.
Care must be taken not to kink or stress the gas
delivery lines.
Pollution Degree: ATD/TD will operate safely
in environments that contain non-conductive
foreign matter up to Pollution Degree 2 in IEC
61010-1.