User`s guide

Installation and Setup
35
(Tenax TA recommended for most applications), gauze
rings and glass wool if you pack tubes or traps yourself.
11. If you are using the LN2 accessory you need to obtain a
supply of liquid nitrogen.
12. Ensure that you have all the required supplies for your
GC and data handling system and that they are up and
running correctly.
13. Review this manual and write down any questions that
arise. When the service engineer is on site he can answer
your questions or provide an alternate source of
information.
Laboratory Requirements
Operating Environment
WARNIN
G
This instrument is not designed for operation in an explosive
atmosphere.
General The site for the instrument must be:
Indoors
Free of dust, smoke and corrosive fumes.
Smooth, level and free from vibration.
Temperature
(safe
operating)
Ambient temperature between 15 and 32 ºC
(59 ºF and 90 ºF). Do not operate the ATD/TD
in a cold room or a refrigerated area.
In addition, the site for the instrument must be:
Free of drafts
Out of direct sunlight.
Away from radiators and heaters.
Humidity
(operating)
20% to 75% relative humidity (non-
condensing)