• Do not use felt tip pens to mark sample tubes as they may
introduce solvents into the tubes. Do not use adhesive labels as
they will jam in the tube handling system.
• Sample tubes recently unloaded from instrument to the carousel
can be very hot. Once the sequence is complete, cool and vent
the sample tubes before you reuse or store them.
• Sample tubes, caps and nuts must be cleaned and conditioned
before reuse. If the sample in the tube is hazardous treat the
PTFE ferrules as hazardous waste.
• After cleaning, carefully check the glass or glass-lined sample
tubes for damage before reusing them. Do not use damaged
tubes. Replace reused sample tubes as necessary. Check for
chips or scratches at the ends of the tube. If the ends are
damaged, the seal may leak and may also damage the o-ring
seals. Also look for glass wool from the retaining plugs at each
end of the tube.
NOTE: The standard TurboMatrix PTFE Analysis Caps
(N6200119) are suitable only for analysis. Swagelock caps
need to be used as storage caps. The TurboMatrix brass
caps with ferrules are suitable for both storage and
analysis.