User`s guide

Troubleshooting
354
Fault: Carousel Motor Stalled
Cause: An obstruction has caused the carousel to halt
rotation.
Action: Turn the ATD off and disconnect the power
cord from the AC mains. Remove the
obstruction and reconnect the line cord. Turn on
the ATD. The instrument will initialize the
motors. It will then return a Ready status if all
of its systems have been initialized correctly.
You must update the tube range on the Run tab
to exclude the tubes that have already been run
and press the Start button to continue your
analyses.
Cause: A mechanical problem has occurred.
Action: Contact your PerkinElmer service engineer.
Fault: Robot Arm Motor Stalled
Cause: An obstruction has caused the robotic arm to halt.
Action: Turn the ATD off and disconnect the power
cord from the AC mains. Remove the
obstruction and reconnect the line cord. Turn on
the ATD. The instrument will initialize the
motors. It will then return a Ready status if all
of its systems have been initialized correctly.
You must update the tube range on the Run tab
to exclude the tubes that have already been run
and press the Start button to continue your
analyses.
Cause: A mechanical problem has occurred.
Action: Contact your PerkinElmer service engineer.
Fault: Sensor Failure Zone 1, 2, 3 or 4
Cause: The sensor, for the sample tube heater, in zone 1 has
failed.
Action: Contact your PerkinElmer service engineer.
Cause: The sensor, for the heated valve and valve block, in
zone 2 has failed.
Action: Contact your PerkinElmer service engineer.