Status Description
SEALER MID
POSITION
Contact your PerkinElmer Service
Representative if you get this message.
SEALER
DOWN
Contact your PerkinElmer Service
Representative if you get this message.
SEALER UP Contact your PerkinElmer Service
Representative if you get this message.
SHOP AIR The shop air pressure is too low (90psi), and
the sensor failed. Adjust the pressure.
SERVICE
ACCESS
Contact your PerkinElmer Service
Representative if you get this message.
TRAP FAILED
LEAK TEST!!
CAREFULLY replace the cracked Cold
Trap, the heated valve may be leaking. Check
the tube for smooth ends, replace O-rings in
the Upper and Lower Seals. If you still get
this message after doing these checks contact
your PerkinElmer Service Representative.
TUBE
HEATER FAIL
Contact your PerkinElmer Service
Representative if you get this message.
TUBE IN
RETURN
POSITION
A new tube has been placed in the active tube
magazine position,
TUBE
JAMMED!
Check for a stuck tube in the current position,
check if the air to robotics has failed. If you
still get this message after doing these checks
contact your PerkinElmer Service
Representative.
TUBE LOAD
FAILURE
Check the position of the pen clip and if there
is anything stuck on the tube. If you still get
this message after doing these checks contact
your PerkinElmer Service Representative.
TUBE
MISSING
Check the magazine for the tube. If you still
get this message after doing this check
contact your PerkinElmer Service
Representative.