Theory of Thermal Desorption
300
Figure 86 Standby Configuration
NOTE: SV is a solenoid valve, and PR is a pressure regulator in
this section.
In standby, flow is maintained through the GC column, and no tube
has been loaded. SV1 is closed. The column head pressure may be
adjusted by the pressure regulator to provide the required gas flow
through the column. SV2 enables carrier gas to be supplied to the
column in the gas chromatograph via the heated transfer line. The
heated valve is in Position B. This isolates the trap from the
atmosphere and prevents the flow of gas through the trap when the
system is idle.