User`s guide
Routine Maintenance
292
To replace fuses:
1. Ensure that the power cord is disconnected from the
power entry module on the rear left side of the
instrument. The fuse drawer is located in the top portion
of this module. Refer to Figure 84.
2. Carefully pry the fuse door open with a flathead
screwdriver.
3. Use the flathead screwdriver to slide out the fuse
drawer.
4. Remove the blown fuse and replace it with a new one of
the correct type and rating. Consult the fuse chart. Both
fuses must have the same rating.
5. Replace the drawer and close the fuse door.
6. Reconnect the line cord.
7. Resume normal instrument operation.
Table 22 ATD/TD Fuses
WARNIN
G
For continued protection against risk of fire, replace only
with the same type and rating of fuse.
Voltage
@ 50/60 Hz
Fuse
(250V rated)
Part Number
100 T 10A M041-7002
120 T 10A M041-7002
220 T 5A M041-7038
230-240 T 5A M041-7038