User`s guide
Routine Maintenance
282
Under normal operating conditions, the filter disks do not require
changing. However, they must be changed if the packing of the cold
trap has been overheated or has moved, or if chromatographic
problems have been observed (such as tailing peaks, sample loss or
sample carryover). You must always replace filter disks when
repacking the trap.
To change the trap filter disks:
1. Disconnect the ATD/TD from the AC line power. Allow
the system to cool. After the instrument has cooled, turn
off the purge gas supply (both carrier gas and dry air).
2. On the TD, open the front door. If you have an ATD,
you need to remove the front plate. Loosen the two
screws on the side of the instrument to release the front
plate and then gently pull the plate up and out.
3. Remove the heated valve cover. You will need a 7/64
Allen key.
4. Using wrenches loosen the fitting on the end of the trap.
Disconnect the fitting from the union and remove the
cold trap.
NOTE: Using wrenches, first loosen the trap nut at the left end of
the trap. Then remove the gas connections at the right end.
Slide the trap to the right to remove it from the peltier
cooler box. Exercise great care. Any torque applied to the
glass tubing could cause it to crack.
CAUTION
Always have a replacement trap available in case of
breakage.
CAUTION
Ensure that the GC oven and the detector are cool. If Ms is
used, the GC-MS transfer line and the Ion source must be
cool. The filament is also off.
WARNIN
G
The heated valve cover and the oven can be hot. Wait for 20
to 30 minutes for them to cool.