User`s guide

Routine Maintenance
251
Figure 70 Location of TD Sample Tube Seals
5. Use the ½” socket supplied in the shipping kit to loosen
and remove the nut on the fixed seal.
6. Use tweezers or forceps to carefully remove the o-rings
and the filter disk.
NOTE: Tweezers or forceps with sharp points may scratch the
inner sealing surface of the nut and cause leaks.
7. It is possible that residue from the o-ring material will
remain on the nut. Clean out any residue in the nut.
8. Using tweezers or forceps, place the smaller o-ring into
the nut. Place the filter disk on top and then lastly place
the larger o-ring into the nut.
NOTE: Do not touch the o-rings or the filter disk with your fingers.
9. Again use the 1/2 inch socket to place the nut onto the
fixed seal assembly. Ensure the nut is threaded properly
and then tighten it until it just seats against the
assembly.
Heated
Valve
Filter
Mobile Seal
Assembly
Fixed Seal
Assembly
O-ring
O-ring
Filter
O-ring
O-ring