User`s guide
Routine Maintenance
247
for the PTFE ferrules (P/N L427-5110) as graphite ferrules are
permeable to air.
Removing and Replacing the Front
Covers of the ATD
You may need to remove the front covers of the ATD in order to
install the transfer line, and perform a number of routine
maintenance tasks. The front covers consist of the arm housing and
the transfer line cover.
1. Make certain that the ATD is in Standby. Turn off the
instrument and disconnect it from the AC line power.
2. If necessary, allow the instrument to cool.
3. Loosen the carousel cap nut and then remove the
carousel.
4. Remove the screw securing the arm housing to the
chassis. This screw is at the bottom of the decapping
station.
5. Loosen the two screws on the left side panel. You do not
have to remove the screws. You must loosen both
screws to loosen the case.
6. Slide the door latch to the left and free the arm housing
from the sides of the instrument and then slide it
outward.
7. You now have access to the tube handling mechanism,
to the heated valve and the cold trap enclosure.
WARNIN
G
The heated valve, the tube oven, and the trap enclosure may
be very hot. Allow 20 -30 minutes for these parts to cool.