User`s guide
Routine Maintenance
244
Table 17 Maintenance Schedule
Cleaning the Instrument
Exterior surfaces may be cleaned with a soft cloth, dampened with a
mild detergent and water solution. Do not use abrasive cleaners or
solvents.
Cleaning and Decontamination
Before using any cleaning or decontamination methods except those
specified by PerkinElmer, check with PerkinElmer that the proposed
method will not damage the equipment.
Item/Operation Frequency
Replace oxygen, moisture and
charcoal filters on the carrier
gas lines.
As required
At least annually
Replace o-rings in the seals As required
Leak test the system As required
Replace the o-rings and filter
disk in the mobile seal.
As required
Replace the cold trap As required
Replace the trap filter disks Whenever the cold trap is
replaced or repacked.
Repack sample tubes Every 2 years or as required
Replace fused silica transfer
line
If required
Clean the carousel area As required