User`s guide

Accessories
184
window.
8. If you are using the Method Editor, save your method. If
you are using the Status tab, you must enter all of the
other method parameters and then save the new method.
9. Run an analysis to test that your devices are connected
properly and your events are programmed correctly.
Example: To turn relay #1 on for 30 seconds, 2 minutes after the
current analysis has started, you would enter: an event time of
120 seconds, Relay Number 1 and Action would be On.
You would now enter a second event: an event time of 150
seconds, Relay Number 1 and Action would be Off.
The following timing diagram shows various devices being
activated, during each analysis. A device can also be activated for
more than one analysis as shown by relay #3.
Figure 56 Timed Event Diagram
Injection
Injection
Injection
Injection
NO NC
ON
ON
ON
OFF
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
1 2 3 4 5 6 min
Start Duratio
n
Relay 1
0.50 1.00
1 2 3 4 5 6 min
Relay 2
0.00 0.05
1 2 3 4 5 6 min
Relay 3
1.50 2.50