User`s guide

Operation
178
3. Enter the desired end tube.
4. From the method drop-down box, select the method to
be used for the selected range of tubes.
5. Press the Add button to add the entry to the sequence.
6. Repeat these steps until you have created the desired
sequence. To delete an entry from the sequence, select
the entry in the list and press the Del button. To change
the tube range, you must add a new entry with the
revised tube range and then delete the old entry.
If you edit the methods called by your sequence, the
sequence will use the revised methods. If you need to
edit a method but do not want to affect your sequence,
then save the revised method with a different name.
NOTE: You will not be able to edit any of the selected methods if
method editing has been disabled from the Preferences tab.
7. Press the green Start button to run the sequence.
The instrument will configure itself based on the first
method. Once the instrument has reached the set points,
the first tube will be loaded and the analyses will
continue as determined by the method.
Shutdown
Normally you do not need to shut down the instrument. You can put
the instrument into economy mode overnight and over the weekend.
See Preferences Tab on page 117. If the instrument will not be used
for more than two days, you may opt to shut it down.
If the instrument is to be disconnected from the GC be sure to cap
the end of the fused silica transfer line using a septum to prevent
contamination of the system.