User`s guide

Operation
164
Figure 48 Config Tab
NOTE: Integrity Testing button is grayed out for all models except
the TurboMatrix 650 ATD. See the section on Integrity
Testing below for operation with a TurboMatrix 650 ATD
PPC Config.
Carrier Gas-Select the type of carrier gas being used. The
correct selection must be made to ensure the proper control of
the pneumatic components.
NOTE: When a change is made to the carrier gas selection, the
power to the instrument must be cycled. This is to ensure
that all of the control systems are reset and purged of the
previous gas prior to using the new gas selection.
Column Length-Input the length and Internal Diameter (ID) of
the column installed in the GC. This information is required to
run the instrument in the Flow Control mode.
Transfer Line-Select the Long or Short radio button for the
length of the transfer line