User`s guide

Operation
159
4. Enter any remaining operators and numbers.
5. Click =.
6. Sqrt calculates the root of the displayed number. X^2
calculates the square of the displayed number.
7. Press the close button (x) on upper right corner of the
title bar of the window to close the calculator.
Reset-Use the Reset command to reset the instrument in the event of
a fatal error or software problem. If the instrument is reset, you must
load your last active method and allow the ATD/TD to reach all of its
set points before beginning an analysis. If you had to reset the ATD
or TD during a sequence, adjust the start and stop tube numbers so
that the sequence includes only the tubes that have not been
analyzed.
NOTE: When you begin the new sequence, the log information
from the previous sequence is lost.
Log Out-The Log Out command closes the tabbed interface and
logs out the current user. The ATD/TD initial screen is displayed.
You can now log on as a different user or power down the ATD/TD.