User`s guide

Operation
155
status of the run. If the sequence has been completed then the time at
which it was completed is also displayed.
Figure 44 Log Tab
The information is overwritten when a new sequence is started.
The Log Report is a log of all events. During a sequence, events can
be logged after every injection, or only deviations may be logged. If
every injection is to be logged, then the Log Report will contain the
following information:
When the method and sequence were started.
When a successful sample injection from a tube is
completed. A line will be appended to the Log Report for
every tube, indicating success or reason for failure. For
example “Tube 1: Analyzed” or “Tube 1: Failed leak test!”
A line will be appended to the Log Report when the method
or sequence is complete.
If the method or sequence that is run was modified but not
saved, the method or sequence name will be followed by the
text “modified”.
If only deviations are to be logged, then the Log Report will only
contain a list of deviations that occurred during the sequence. It will