User`s guide
Operation
143
Figure 38 Internal Standard Popup
You will need to enter a standard injection time by selecting the
Preferences entry on the Tools tab and then selecting the Internal
Standard button, from the Config tab.
This calls up a dialog box that allows you to enter the Inject Time as
well as the Loop Load and Loop Equil for the internal standard.
Enter an injection time between 0.1 and 9.9 minutes.
Enter a Loop Load time. This will determine the time interval that
ISV1 is open to load the sample loop in preparation to introducing
the Internal Sample onto the sample tube. The length of time
required is a function of the loop load flow rate and the volume of
the loop. Typically, 10 loop volumes should be passed through the
loop to ensure complete filling of the loop with the standard.
10XLoop Volume (mL) / Load Flow (mL/Min) = Time (Min)
NOTE: Refer to the Accessories section of this manual for
information on setting the Loop Flow Rate, Tube Flow Rate
and the Loop Equil time.