User`s guide
Operation
139
5. To save your settings as a new method or to update the
existing method, press the Tools button and select Save
As.
6. To save the method existing name press OK. To create a
new method, enter a new name for the method by
pressing the method name. An alphabetic keypad is
displayed. To enter numeric characters, press the Num
key. Press Alpha to return to the alphabetic keypad.
7. Press OK to enter the new method name and then press
OK save the method and return to the Status tab.
Timing Tab
There are four timing parameters required to complete an ATD/TD
method. The Purge time, primary Desorb time, the Trap time and
the Cycle Time. Additionally, if you have the Internal Standard
Addition Accessory installed and enabled, you will need to enter a
standard Inject Time. The exact values that you enter here will
again be based on your application. The ATD/TD timing parameters
will also be affected by the GC method.