NOTE: Before powering up the TD/ATD again, ensure that
a) Carrier Gas supply and Dry Air to the TD are restored,
b) Wait 20 to 30 minutes to ensure that both the cold trap
and its environment are absolutely dry before power is
applied to the TD.
Programmable Pneumatics Control
(PPC) TurboMatrix Series 300 TD,
350ATD and 650 ATD
System Setup
The PPC module is being configured and the initial pressure or flow
setpoint is being established during the SYSTEM SETUP step. This
setup happens automatically. The touch screen will display
SYSTEM SETUP in the status line (see the following screen).
At instrument startup do not change any parameters in the method,
the configuration, the download methods or the configuration from
the remote control software until the instrument has successfully
completed the SYSTEM SETUP step.
If the SYSTEM SETUP step fails, the instrument must be power
cycled. A common cause of this will be no carrier gas.
Status Line