User`s guide
Operation
130
The method can be stopped at any time by pressing the
Stop button. A confirmation pop-up window will
display, press Ye s to confirm that you want to stop the
run. The analysis on the current tube is aborted and the
tube will be unloaded and returned to the carousel. The
method will be interrupted and the instrument will revert
to Standby.
Multiple Method (Sequence) Operation
You can create a sequence of methods if you have varied
applications. Multiple Method Operation must be enabled on the
Preferences/Run tab. You must also have created and tested all of the
methods required for your sequence.
Your ATD application may require the use of multiple methods for
analysis of the 50 tubes. It is also possible to run a sequence of
methods on a single tube loaded into a TD, if required.
1. Select the Sequence option from the Run tab.
2. Press the start tube entry box and then press the plus or
minus button to select the desired start tube.
3. Enter the desired end tube.
4. From the method drop-down box, select the method to
be used for the selected range of tubes.
5. Press the Add button to add the entry to the sequence.
6. Repeat these steps until you have created the desired
sequence. To delete an entry from the sequence, select
the entry in the list and press the Del button. To change
the tube range, you must add a new entry with the
revised tube range and then delete the old entry.
If you edit the methods called by your sequence, the
sequence will use the revised methods. If you need to
edit a method but do not want to affect your sequence,
then save the revised method with a different name.
NOTE: You will not be able to edit any of the selected methods if
method editing has been disabled from the Preferences tab.