User`s guide
Operation
127
During primary desorption, the carrier gas will flow through the
sample tube from the bottom to the top. It is therefore essential that
sample tubes are loaded with their sampling ends (grooved end
)
upwards
. This will ensure that the volatiles are desorbed from the
collection end of the sample tube and do not pass through the whole
length of the adsorbent bed.
Sample tubes with pen clips may be loaded onto the TD. The tube
must be placed into the TD with the pen clip facing back, toward the
back of the instrument.
To load the tube:
1. Remove both caps from the tube. Ensure that the
instrument is Ready.
2. Place the tube into the tube compartment with the
sampling end (grooved end) upwards.
3. Center the tube in the lower, mobile seal and push down
on the seal with the tube. Gently push the tube in and
center the tube below the upper seal while it is seated
and then release the tube.
4. Press the Seal Up button to raise the tube into the
mobile seal.
5. Press Start.
6. Once the tube has been sampled, the display status will
be: Cooling. Do not touch the tube while this message is
displayed. The tube will still be hot from the desorption
process.
7. When the instrument status changes to Ready, press the
Seal Dwn button. The Mobile Seal with the tube will
move down and the tube can be released.
WARNIN
G
Ensure that sample tubes are cool before handling them.