User`s guide

Operation
126
groove, and prying them apart. Other techniques, such as twisting
the pen-clip from the tube, can cause permanent damage to the
pen-clip. Do not use pen-clips that have been permanently deformed
(loose).
Up to 50 capped sample tubes may be loaded onto the carousel. The
sample tube should be loaded onto the carousel with the
TurboMatrix brass analytical caps in place on both ends of the tube
or the PTFE analysis caps. The ATD will remove the caps when the
sample tube reaches the analysis position, and replace the caps after
the tube has been desorbed.
The front 15 positions of the carousel are available for loading. The
other positions may be accessed when the instrument is in the
Standby state, or when the instrument has been paused. Press the
Rotate button on the display. Do not rotate the tray by hand.
To load the ATD carousel while on the instrument:
1. Install the tubes on the carousel in the positions 1 to 50.
2. By pressing the Rotate button, you can move the
carousel one quadrant at a time to gain access to tubes.
Loading a Single Tube into the TD
WARNIN
G
Recently unloaded tubes can be very hot. Ensure that sample
tubes are cool before handling them
WARNIN
G
Never try to access a tube on the TD while it is running. The
oven heater may be activated and can cause a serious burn