User`s guide
Operation
123
1. Select the Setup tab.
2. Press the Economy Mode check box. A check mark
will appear in the box to enable the option.
3. Press the Config button to open the configuration dialog
box.
Figure 32 Economy Mode Dialog Box
4. The instrument will only go into Economy mode during
a specified interval. Select the time during which the
instrument is allowed to enter Economy mode.
5. The ATD/TD will only go into economy mode after a
predetermined interval during which there is no activity.
You can select this time. If you select 0, then the
instrument will go into economy mode at the beginning
of the selected interval, providing the analysis is not in
progress. If you enter 30 minutes, then the instrument
will go into economy mode 30 minutes after all analyses
have been completed.
6. Now enter the time for the ATD/TD to “wake up”. At
the “wake up” time, the ATD/TD will switch on the
heaters and re-activate the purge valve. The instrument
will configure itself according to the method specified
on the Run tab.
7. Once you have entered the desired settings, press OK to
accept the settings and close the tab.
8. Confirm that the date and time are correct.