8.3.9.5 Reset Recovery Time Multiplier (default 0)
Multiplies the Reset Recovery Time, including the recovery time of the field reset
command.
Reset Recovery Time Multiplier Reset Recovery Time
0 1x
1 2x
2 3x
3 4x
Figure 8-15: Reset Recovery Time Multiplier
8.3.9.6 Enable ISO14443 B Anti-collision (default 0)
If set, the anti-collision algorithm for ISO 14443 B tags is enabled.
8.3.9.7 Disable ISO 14443-4 Error Handling (default 0)
If set, ISO14443-4 Error Handling is disabled. The error handling always uses the
TMR time-out.
8.3.10 Reset Off Time (14h)
The Reset Off Time register represents the field off time in ms.
This register is used for the select, continuous read and multi-tag commands.
Default value is 0Ah.
8.3.11 Reset Recovery Time (15h)
The Reset Recovery Time register represents the recovery time in ms after the field
is turned on.
This register is used for the select, continuous read and multi-tag commands.
Default value is 25h.
8.3.12 Application Family Identifier (16h)
The AFI (Application Family Identifier) is only supported for ISO14443B tags. If the
set value is different from 00h, the AFI is implemented in the ReqB command of
ISO14443B. Only transponders with an identical AFI will answer to the reader.
Default value is 00h.