ATE-100 HAND HELD CALIBRATOR OPERATING MANUAL REVISION 4.
Thank you for your purchase. Please feel free to contact us if we may be of assistance in any way. Should you require service for this instrument or wish to take advantage of our expert factory recalibration services, please contact the factory for return information: Ashcroft Inc. 250 E. Main Street Stratford, CT 06614-5145 Tel: (203) 378-8281 Fax: (203) 385-0402 For the latest product information, please visit our website at: www.ahscroft.
Table of Contents Section Description Page Introduction .................................................................................................................................. 4 Base Unit Overview......................................................................................................... Quick Select Module Overview ...................................................................................... The GQS-1 Quick Select Pressure Module........................................
Section Description 12.0 Damping Function . ...................................................................................................................... 20 12.1 12.2 Set Up of Damping Function........................................................................................... Activating or Discontinuing Damping............................................................................. 20 21 13.0 Percent Function ....................................................................
Section Description Page 21.0 Dual Module Functions . .............................................................................................................. 42 21.1 21.2 21.3 Accuracy of Dual Module Measurements........................................................................ Setting Up Dual Module Differential Pressure Measurement.......................................... Setting Up a Dual Module Summation Pressure Measurement....................................... 42 42 44 22.
Section 1.0 units can be independently selected from the library of twelve factory programmed or one operator designated engineering unit. Temperature measurement data can be displayed in degrees Celsius, Fahrenheit, Kelvin or Rankine. In addition, when used with the GQS-RT1 Quick Select module for temperature measurement with an RTD temperature probe, the calibrator system can display the RTD measurement in ohms.
in the Quick Select module. As such, any Quick Select module can be used in any base unit and the measurement system will provide measurement accuracy in conformance with the published specification. Section 1.4.1 micro-machined piezoresistive strain gauge sensor. This technology takes advantage of the fact that, when put under stress, (as with the flexing of a diaphragm under pressure or vacuum) the resistive properties of a piece of silicon will change.
programmed into the GQS-RT unit to provide enhanced accuracy. Each GQS-RT interface module can accommodate up to 8 different programmed RTD calibration curves. Programming of the RT module is accomplished via an optional software package and any PC compatible computer with an available standard serial communication port. See Appendix E page 75 for probe connector pin out information. Section 3.
opening, starting closest to the battery compartment and reattaching Velcro. The strap may be adjusted in this fashion to provide for proper strap length to accommodate any hand size. Section 3.2 base unit should contain a Quick Select module to protect from dirt or other debris getting into the base unit assembly. If only one Quick Select pressure module is needed install the System Protection Module (part number GQS-XS) supplied with your unit at the time of shipment when using the HHC system.
where: Quick Select pressure module(s) to be used to the pressure source to be measured. If a gauge pressure measurement is to be made using a differential Quick Select pressure module be sure to connect the pressure to be measured to the high pressure port on the Quick Select pressure module.
BatFunc UserEng PerCent H2Oref ON OFF Step 2 With the HHC on and reading pressure press the Step 3 With “BatFunc” flashing press the enter key. battery check key, designated BAT CK. ENT SET UP Step 4 Using the up/down arrow keys to select Auto Shut off. When selected the text will flash. Step 3 Using the arrow keys select the battery save Auto shut off CfgBattIndicator function (designated BatFunc in the set up menu). When selected, the text “BatFunc” will flash on the display.
battery voltage level when the BATCHK key is pressed. Section 3.4.3 DateTime will flash on the display when selected. Owner Alarm DateTime RS232 Low Battery Icon The HHC provides advanced warning of a low battery condition. The icon, which looks like a miniature battery will appear and flash in the center of the top line of the display when the measured battery voltage drops to 5.7 Vdc. The calibrator will continue to function properly with the low battery icon flashing.
initial setup or when the lithium battery used to back up the RAM needs replacement (every 1-2 years depending on the environmental conditions under which the HHC is used). RAM memory is only supplied if unit is purchased with data logging option. Time/date information is not maintained in HHC units without data log option. Section 3.4.5 information. Step 6 With the proper access code displayed press the enter key.
been entered and is displayed press the enter key. ENT The HHC will respond: JOHN Q OPERATOR Ent=OK CE=Cancel Step 10 To store the owner/user information displayed Designation Function On/Off Set Up Turn unit on and off. PORT SEL Used to select displayed measure-ment data. Selection includes: reading pressure from either one or both installed sensors as well as selection of reading pressure and electrical measurements simultaneously. (See Section 5.0 for details.
Designation Complete details of each of the above outlined capabilities is included in the section referenced after the function summary provided above. Function DATA STORE Provides ability to store measured pressure or temperature values. (See Section 22.0 for details.) HOLD Freezes the displayed pressure and electrical measurements. (See Section 10.0 for details.) ^ Moves cursor up. *TRIP DETECT Provides access to pressure < *DAMP Moves cursor to the left. > CE mA/V v *% ENT Section 5.
the display returning to a readout mode featuring the measurement data from both of the installed modules. As seen below: F +72.35 o psi mA +00000 +1234 Step 3 To resume the display of pressure measurement psi +123456 data, for the pressure module not zeroed, press the port select key the required number of times (one or two) to restore the dual pressure display. See 8.4 for display of pressure in two different engineering units. When complete the HHC will display: See Section 7.
Result from pressing mA/V key a second time: The engineering unit abbreviations correspond to the following engineering units: psi volts +123456 +1.
the steps below. down arrow keys to increment or decrement through the alphabet/numerical values and the left and right arrows to move the active entry field (cursor) to the next location. Step 1 With the HHC on and reading pressure press the key designated SET UP. SET UP If needed for the desired engineering unit, a blank can be inserted by using the down arrow with the letter A flashing on the display.
ENG UNIT Step 1 With the HHC on and measuring pressure press the engineering unit key. Designated ENG UNIT on the keypad. Step 2 Using the arrow keys select either of the two ENG UNIT desired engineering units. For this example we will select psi as one of the desired units.
The HHC will respond by displaying: psi Bar inHg mBar most commonly used by manufacturers of pressure instruments when calibrating pressure measurement devices in terms of the height of a column of water. To facilitate your calibration and test of these instruments, the HHC includes 3 conversion factors for each of the water column based pressure engineering units. The conversion factors are for the temperatures of 4 and 20 degrees Celsius and 60 degrees Fahrenheit.
Section 9.0 Warning: Failure to enter the zero (0) will result in the Tare Function HHC automatically accepting the last displayed value as the desired tare value. The HHC has the ability to tare (subtract an operator selected value) from the displayed pressure or temperature measurement value of either one or two installed Quick Select modules.
Step 2 To exit the hold function press any key. not activated the HHC takes a new pressure value every 130 to 240mS (~8 times per second) and updates the display. If required, this same measurement data is also transmitted over the RS232 interface. Note: While in the hold mode, if the RS232 interface is enabled, the HHC will continue to transmit the displayed values. Section 11.0 When damping is activated the value displayed and/or transmitted is an average value.
down results from the fact that an averaged value will be displayed. If the damping level is set to 16 the displayed value will represent the average pressure measurement over the previous ~2 seconds (16 readings times 130 mS/reading). Therefore, the greater the level of damping in use the greater the time lag will be between the displayed measurement and the true (not averaged) pressure measurement value. Section 13.
ENT Step 7 With the desired output “zero” keyed in and displayed press the enter key. The HHC will respond by displaying: ENT Enter Input Zero .000000 The HHC will respond by displaying: Enter output FS 0.0000 Step 4 Using the number keys enter the lowest calibra- tion point for the device to be tested. For example: Step 8 Using the number keys enter the full scale output of the device being calibrated.
of operation. Paa = the absolute pressure in duct in psia T = air temperature in duct in degrees Fahrenheit Flow Volume = (velocity x duct area)/144 % Step 2 Connect the output lines from the device under 2 – Purchase of a characterized and certified pitot tube test to the miniature banana jack inputs on the front of the HHC. Use the mA and COM jacks for current devices and the V and COM jacks for voltage output devices. or annubar with a K factor determined for the specific probe to be used.
Step 3 With the word Velocity flashing press the These are the 4 units of measure available for display of flow velocity. enter key: ENT Step 2 Using the left/right and up/down arrow keys select the desired unit of measure. When selected the unit of measure will flash. In this example the engineering unit of ft/sec has been selected. The HHC will respond by displaying: Enter Air Density lbs/ft3 .
W = width of the duct in inches H = height of the duct in inches Step 3 With the desired unit of measure flashing, press the enter key. ENT For round ducts: area (square inches) = (π/4 )D2 After pressing the enter key the HHC will resume measuring flow volume in the selected unit of measure. π = 3.141592 Section 14.
MeterSec +12345 Step 4 Next, press the port select to deactivate the display of flow measurement data on the side that you wish to maintain displayed measurement data in terms of ft/min. A pressure display is deactivated when it is replaced by a milliamp (mA), Voltage (V) or a blank (------). mA +1234 Step 8 Reactivate the right side of the display for flow measurement by pressing the Port Select key, as required. See Section 5.0 for details on the port select function.
UNIT key. Details on engineering unit selection are provided in Section 8.0 of this manual. Quick Select module. For example, the measurement of flow can be displayed in ft/min for one installed Quick Select pressure module and the second installed pressure module can be used to simultaneously display the temperature measurement value in degrees Celsius or Fahrenheit. This ability eliminates the need for “on-thefly” change between flow and temperature measurement functions when doing airflow testing.
module. This ability eliminates the need for “on-the-fly” change between flow volume and flow velocity measurement functions when doing airflow testing. Step 4 Using the left/right arrow keys select the desired measurement function. In this case we have decided that the left side of the display will remain in the velocity mode and the right side will be changed to measure flow in terms of volume. Therefore, volume should be selected for this example.
measurement (velocity or volume) mode of operation. Refer to Section 14.2 for information on setting up the HHC for velocity measurement or Section 14.4 for information on setting up the HHC for volume measurement. leak rate mode the HHC will monitor the pressure vessel over the programmed time interval and display the leak rate in terms of cc/seconds. Use of the leak detection function requires the setup of the leak test parameters.
Step 9 Using the number keys enter the time inter- Step 8 With the desired test interval displayed press val, in seconds, over which the leak rate test is to be performed. The time interval may be from 1 to 65,000 seconds. the enter key. ENT At this point the HHC will return to basic pressure measurement. Time to Monitor sec. 12345 See Sections 15.3 and 15.4 for information on performing leak rate testing. Step 10 With the desired test interval displayed press the enter key. Section 15.
of the HHC. --- “ on the other side of the display. If a second module is installed in the other sensor bay it can be re-selected using the port select function (PORT SEL) as detailed in Section 5.0. LEAK The HHC will display: Timing eng unit sec xxxxx Section 16.0 The HHC can be used with most standard RTDs (Resistance Temperature Detectors) to provide precision temperature measurements. An GQS-RT type Quick Select module is needed in order for the HHC to perform temperature measurement with an RTD.
ON OFF made in degrees Celsius. Engnring units: C oF K oR o Step 4 To access the temperature setup menu press the Set up key. Step 11 With the desired engineering unit selected and SET UP flashing press the enter key.
Probe: 12345678 P&a385 RO = 100 Resolution: 1 .1 01 0.001 Step 13 With the desired probe type selected and flash- Step 15 With the desired resolution selected and flash- ing press the enter key. ing press the enter key. ENT ENT The calibrator will respond by displaying the following: Resolution: 1 .1 .01 The calibrator will briefly show internal limit settings and will then commence temperature measurement. 0.
Section 17.1 the millivolt mode has been selected, the programming cycle will end when the “enter” key is pressed, and the calibrator will begin to display the real-time millivolt measurement from the thermocouple circuit. Thermocouple Measurement Set Up Step 1 To begin measurement with a thermocouple, first attatch the thermocouple to a male “miniature thermocouple connector.” Please note that connectors are specifically manufactured for each thermocouple type and will be marked accordingly.
deactivate (convert the display to mA, “------” or Volts) the other side of the display. the reading from the other module bay, if a module or an interface module has been installed. PORT SEL If the “manual” mode was selected, the calibrator must be programmed with the reference temperature of the external reference junction. the display will read” Ref Step 2 Press the mA/V key as required to change the temperature electrical measurement parameter to that to be used for the switch test.
change of state has been detected. When the switch has tripped the HHC will display: Step 3 With the word “Deadband” flashing press the enter key. Eng Unit Tripped +12345 +vvvvv ENT Step 4 Gradually increase or decrease, as required, pressure, vacuum or temperature until the display of the HHC indicates that a change of state has been detected.
Disable Manual Deadband 6 7 8 9 Step 11 With the word “Disable” flashing press the enter key. Section 19.1 ENT The switch deadband test operation has now been terminated. Section 19.
Owner Alarm DateTime RS232 +0.003 in.H2O +0.0000 psi Left Min: +0.0033 Max: +0.0035 Right Min: +0.0000 Max: +0.0000 Step 3 With the text RS232 flashing press the enter key. Each data set consists of three (3) lines. The first line displays the current measurement data as seen on the instrument display. This line will display the pressure, flow, temperature or electrical measurement in the selected units of measure.
Disable ISO1745 Journal Inquiry sion of data during the setup process. For most functions data transmission will resume when the set up is complete. However, the RS232 interface must be re-enabled via the RS232 set up process to re-initiate data transmission after either automatic data logging or leak detection functions are activated. Step 10 With the word Journal flashing press the enter key.
Disable ISO1745 Journal Inquiry move through the menu selection and highlight the text “RS232” so that is flashes on the display. Owner Alarm Date Time RS232 Step 11 With the word Inquiry flashing press the enter key. ENT Step 3 With the text “RS232” flashing press the enter key. ENT The HHC will respond by displaying the following: Inquiry Char. xx.x This will activate the RS232 setup options sub menu.
sion of data during the setup process. For most functions data transmission will resume when the set up is complete. However, the RS232 interface must be re-enabled in the RS232 set up process to re-initiate data transmission after either automatic data logging or leak detection functions are activated. Section 20.0 where: dd/dd/dd: the date of the last calibration of the left module. Type: provides information on the type of Quick Select pressure module.
2 – The percent life remaining is based on the character- capability is provided in Section 21.2 entitled Dual Sensor Differential Pressure Measurement. istics of standard 9 Vdc alkaline batteries. Use of other batteries types, such as; zinc, lithium or rechargeable NiCad batteries, will result in inaccurate estimation of remaining battery life in both percent indication and bar graph indication.
Step 2 With the modules installed, turn the calibrator When the dual sensor differential firmware is active the following measurement sequence will result from pressing the PORT SELECT (PORT SEL) key. on by pressing the on/off key. ON OFF First press of the port select key: L-R PSI xxxx Step 3 With the calibrator on and displaying pressure measurement data from the two installed Quick Select modules press the Set Up key.
Section 21.3 tion press the port select key one time after the set up has been complete. The calibrator will display the following text Setting Up Dual Module Summation Pressure Measurement Step 1 With the calibrator off, install the two Quick L+R PSI PSI xxxx xxxx Select pressure modules in the calibrator. As this is a summation process the modules may be in either sensor bay. for a calibrator displaying measurement values in the psi engineering unit.
the L+R indication will be not be present on the display. The display is limited to 16 characters/line and cannot display both the L+R and a U_XXXXXX engineering unit label. Section 22.0 When used in conjunction with the Event Timer function (see Section 24) the HHC system can be set up to take data during a selected future time period at an operator programmed time interval.
Section 22.3 Section 22.4 Labeling Function – Standard Data Logging Mode Automatic Data Logging – Standard Data Logging Mode The label option allows the input of an alphanumeric label of up to 16 characters in length. Multiple labels can be used. For example, a label can be entered and data taken to be stored under the label.
Select pressure modules to be monitored and logged. Section 22.5 Readings to Log None Left Rght Both Manual Data Logging – Standard Data Logging Mode Follow the procedure below to conduct manual, operator controlled, data logging. In this example we have opted to log the output from both Quick Select pressure modules. Step 1 With the HHC system on and reading pressure press the data log key.
desired storage format. The desired format will flash on the display when selected. press the data log key on the keypad. DATA LOG Date/Time Stamp Disable Enable The HHC system will respond by displaying the following: In this example we have opted to forego the date/time stamp of the stored data values. THE LAST DATA LOG MENU SCREEN USED Step 6 With the desired format selected and flashing Step 2 Using the up/down and right/left arrows press the enter key.
> ^ v % mA/V brings the next highest stored address value to the display. skips 10 stored values in increasing address direction. skips 10 stored values in decreasing address direction. brings first data point to display. brings last data point to display. Step 3 With the word review flashing press the enter key. ENT Step 4 Data points may be browsed through by using the keys as indicated below: Key brings the next lowest stored address value to the display.
Step 2 Using the up/down and right/left arrows browse Section 23.1 through the data log menu and locate the word “Erase”. The menu line with the word Erase appears as: Label Step 1 With the HHC on and displaying measurement Erase data press the data log key. DATA LOG Step 3 Using the left/right arrow keys select the word Erase. When selected the word Erase will flash. Label Erase The HHC will respond to the data log entry by displaying the following.
Step 3 Using the arrow keys, highlight the text “CertGen Use the up and down arrow keys to enter letters and or numbers to define an instrument tag name. The entry/ selection sequence is as follows: to proceed with the certification generation process ”. CertGen Exit Erase Up/down arrows provide access as follows: 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, _, A, B, C, D, E, F, G, H, I, J, K, L, M, N, O, P, Q, R, S, T, U, V, W, X, Y, Z Step 4 With the text “CertGen” flashing on the display press the enter key.
The HHC will respond by prompting the operator to enter the number of “as found” data points to be taken. A data point consists of the pressure measurement as read on the calibrator and the current or voltage output from the device under test as measured on the calibrator. Therefore, each data point is actually two (2) measurement values. The prompt appears as follows. operation.
DATA STORE Step 15 Note: If the HHC has an GQS-RT or TC temperature module and the probe is installed it will automatically offer the opportunity to store the measured temperature value. After the last set of data to be logged has been stored the HHC will prompt the operator with the following display. AsFound readings View Done Redo If a different value is to be entered or if the HHC is not equipped with a temperature module the temperature can be entered via the numeral keys on the keypad.
STORE ment being added to the calibration record for the device under test: After the last set of data to be logged has been stored the HHC will prompt the operator with the following display. “Use AF for AL” AsLeft readings View Done Redo Step 18 To take “As Left” data and store the measured values in the HHC’s memory use the arrow key to highlight the text “yes” under the “As Left” data prompt screen.
Note: If the HHC has a GQS-RT or GQS-TC1 tempera- New Tag: will result in initiating a new calibration certification process. Previous Tag: will allow the review of data previously taken for specific tag names. ture module installed, it will automatically offer the opportunity to store the measured value the base unit is receiving from the RTD or thermocouple.
Step 9 With the desired baud rate flashing press the Follow the procedures below to change the battery back up unit. enter key. ENT Note: The HHC system should be turned off prior to initiating battery replacement procedure. The HHC system will respond by displaying: Step 1 Turn the HHC indicator bottom side up and Disable ISO1745 Journal Inquiry locate the four (4) phillips head screws that secure the upper and lower case halves together.
HHMMSS XXXXXX isfy this requirement. Delayed initiation of unattended measurement activities requires the set up of both the Event Timer and data logging function. Section 25.1 Step 6 Enter the desired stop time using the number keys. When properly entered a stop time of 11:30 PM will look as follows: Event Timer Set-Up HHMMSS 233000 To program the HHC system for delayed monitoring and data logging follow the steps below. appears on the display of the HHC system press the enter key.
Enter Interval sec. xxxxxx age format. The desired format will flash on the display when selected. Date/Time Stamp Disable Enable Step 4 Using the number keys enter the desired time interval in seconds. The interval may be from 0.1 seconds to 65,000 seconds. In the above example the date time stamp function has been enabled and the stored data will be labeled with time date information in the database. Enter Interval sec.
High Setpoint ???????? Important Note: Set points must be entered for the spe- cific engineering unit to be monitored. For example if a setpoint of 10 is entered the alarm relay will trip when measured and displayed pressure exceeds the numerical value of 10. The HHC will not automatically convert the entered trip point to a new value to reflect a change in engineering units. Open Close Pin 8 Relay 2 common Pin 9 Relay 2 (normally opened or normally closed) Section 26.
Section 27.0 Recertification of the Calibrator Section 27.3 Both the base unit and the pressure modules can be recalibrated and recertified in the field. The base unit includes a firmware based, menu driven procedure that provides operator prompts to facilitate the recalibration process. The procedure encompasses both the acquisition of “as found” data and performing adjustment and recalibration of the base unit and modules. The recertification program is accessed via the set up menu key.
accompanied the product shipment from the factory. See section 27.0 “Access Code” for information on lost passwords. connections for the input of the voltage standard to the calibrator base unit. To check the measured voltage using a precision voltage source apply 1.00000 volt ±50 microvolts. Step 12 Step 8 With the correct access code displayed press the enter key. Do not use a voltage less than or greater than 1 volt ± 50 microvolts.
To apply the 0.1 Vdc signal use the same procedure as outlined in steps 10 and 11 of this section. vided with the calibrator base unit: Access code? 12345 Step 15 The value displayed under the dashed line rep- resents the as received reading for the .1 Vdc measurement. The reading will look something like follows: Step 7 With the correct access code displayed press the enter key. ——–––— 0.1V +.099998 RefCheck ENT The calibrator will respond by displaying: Take note of the measured value.
After the calibrator completes auto-adjusting the measured value it will display: Apply 10.0000V to volts jack Pin 6 Step 14 Using a precision voltage generator apply a 10.0000 volt (±500 microvolts) input to the miniature recessed banana jacks. Use the COM and V jacks on the calibrator keypad. Pin 7 Step 15 When the proper voltage is being applied press *****WARNING***** the enter key.
Calibration done Save Cancel of the calibrator system is not degraded by the calibration process. As recommended by standards organizations such as A.S.M.E., N.I.S.T and various U.S. and world government agencies calibration standards should be at least 4 times more accurate than the device to be calibrated. Step 20 Using the left/right arrow keys select “Save” to write the new data to the EEPROM calibration storage. To discard the electrical readings select Cancel.
Step 8 Enter the 5 digit calibration access code pro- Select pressure module and the flashing word “working” indicates that the calibrator is making the internal adjustments necessary for the recalibration. vided with the calibrator base unit. Access code ? 12345 After approximately 3 seconds the calibrator will display: Step 9 With the correct access code displayed press Full Scale Cal Yes No the enter key.
Linearity Cal Yes No linearization mode. The zero and span values established in these procedures will be used for the zero and span values of the new calibration data. If zero and span are not reset at this time the recalibration will use the values for zero and span stored in memory from the previous zero and span activities. Step 18 If calibration activity is to be limited to zero and span, using the arrow keys select the word No (when selected the word “No” will flash on the display).
Calibration done Save Cancel Step 31 With the applied pressure displayed press the enter key. ENT Step 36 To save the new linearization data to memory, thereby overwriting the previously stored values in the EEPROM, select “Save” and press enter. To discard the newly generated calibration information select “Cancel” and press enter. The calibrator will respond by displaying: +xxxxx working After 3-5 seconds the calibrator will request the next calibration point by displaying.
TO CONVERT FROM TO Revision 4.6 3/2004 29.8368 1.33322 33.8639 .980665 980.665 2.55851 30.702 100.731 0.0193368 0.491154 .0142233 14.2233 .037108 0.4453 1.46096 mmHg @ 0°C in.Hg @ 0°C gm/cm1 kg/cm2 in. sea water ft sea water m sea water 0.97891 0.014198 cmH2O @ 20°C 0.432756 10.0 0.145038 kPa ftH2O @ 20°C 0.01 1.45038 10-4 Pa 2.4864 1013.25 14.6959 atm 0.036063 1000 14.5038 bar in.H2O @ 20°C 1 0.0145038 mbar mbar 68.9476 psi 1 psi atm - 70.100730 .
APPENDIX B – ASCII CHARACTER CODE This chart shows the ASCII character set and corresponding code numbers in decimal, hexadecimal and binary form.
Decimal 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127 128 129 130 131 132 133 134 135 136 137 138 139 140 141 142 143 144 145 146 147 148 149 150 151 152 153 154 155 156 157 158 159 160 161 162 163 164 165 166 167 Revision 4.
Decimal Binary Hexadecimal 176 177 178 179 180 181 182 183 184 185 186 187 188 189 190 191 192 193 194 195 196 197 198 199 200 201 202 203 204 205 206 207 208 209 210 211 212 213 214 215 216 217 218 219 220 221 222 10110000 10110001 10110010 10110011 10110100 10110101 10110110 10110111 10111000 10111001 10111010 10111011 10111100 10111101 10111110 10111111 11000000 11000001 11000010 11000011 11000100 11000101 11000110 11000111 11001000 11001001 11001010 11001011 11001100 11001101 11001110 11001111 11010
APPENDIX C - PRODUCT SPECIFICATIONS* BASE UNIT PHYSICAL SPECIFICATIONS PRESSURE SENSOR MODULE SPECIFICATIONS Dimensions 7.88 in. (L) x 4.24 in. (W) x 3.25 in. (H) GQS-1 Weight Max. 2.2 lbs. w/2 pressure modules installed Case Material High impact ABS Sensor Module Capacity 2 bays for GQS “Quick Select®” sensor modules Display 2 line LCD, 0.037 in. height per line.
PRODUCT SPECIFICATIONS* STANDARD RANGES GQS-2 psi (gauge and absolute pressure) *5 10 15 20 25 30 50 60 100 150 200 250 300 500 600 1000 1500 2000 2500 3000 5000 6000 7500 10,000 GQS-1 ˝H2O (gauge/ differential pressure) 0.25* 0.5* 1.0* 2.0* 3.0* 5.0* 10* 15* 25* 50* 100* 150* 200* DIMENSIONS psi in. H2O in. Hg ftSW Bar mBar kPa mPa mmHg cmH2O mmH2O kg/cm2 User Selectable 3.25 4.24 .43/.58 Other Engineering Units** PTE-1 7.
SYSTEM ACCURACY WITH GQS-TC1 THERMOCOUPLE INTERFACE Table I System Accuracy (reading in temperature units): Includes the HHC base unit and GQS-TC1 interface module. (Does not include inaccuracy of the thermocouple device. Consult thermocouple manufacturer or ISA MC96.1 for thermocouple accuracy specifications. Typical inaccuracies range from ±1 to ±2.2°C.
APPENDIX D RTD PROBE CONNECTOR SWITCHCRAFT TINI QG (TA4F) PIN 1 2 3 4 FUNCTION 2W+ 4W+ 4W– 2W– 2 WIRE RTD TA4F 1 2 3 4 3 WIRE RTD TA4F 1 2 3 4 4 WIRE RTD TA4F 1 2 3 4 - 77 - Revision 4.
APPENDIX E Certification of the TC1 Thermocouple Interface Module Theory With the TC1 module installed in the base unit (set-up to display in mV), a precision millivolt calibrator/supply will apply a known mV signal to the module/base unit system (simulating the Seebeck voltage) in place of a thermocouple. The applied mV signal can then be read on the base unit display and compared to the nominal input provided by the precision millivolt calibrator/supply. The TC1 module cannot be adjusted.
- 79 - Revision 4.
Visit our web site www.ashcroft.com All specifications are subject to change without notice. All sales subject to standard terms and conditions. © Ashcroft Inc. 7/01 I&M002-10037-11/00 (AHHC-MAN) .5CDPS3/04 1P6/00 Revision 4.6 3/2004 - 80- Ashcroft Inc. 250 E.