User's Manual

TD 93021US Rev A
10 FEB 2014 / Ver A 72
Installation Guide
teleCARE IP
The image below shows the location of the DIP switch which is in the same place on all active switch
modules.
Figure 89. Active switch module DIP switch location
The the DIP switch allows 4 addresses to be set: 0, 1, 2 and 3. These settings are only used for active
switch modules and active socket modules.
Address 4 is used for the toilet cancel module and this module has the address fixed and it is not set
by a DIP switch.
The addresses 5 to 7 are used for special teleCARE IP peripherals consisting of the corridor lamp and
the duty selector. These module have the address fixed and it is not set by a DIP switch. Addresses 6
and 7 are reserved for future extensions.
The room bus address assignments are summarized in the following table:
6.5 Bedside Module (NIBM2)
The Bedside Module (NIBM2) is designed for use in the teleCARE IP system. It is suitable for use with
all teleCARE IP handsets and it supports teleCARE IP speech and entertainment.
Room Bus Address Active Peripheral Address Setting
0 - 3
Door side Modules
Bedside Modules
Pull Cord Modules
Medical Rail Socket
Set by DIP switch
4
Toilet Cancel Module
Fixed
5
Corridor Lamp
Duty Selector
Fixed
6
Card reader
Fixed
7
Room Display
Fixed
DIP Switch
Settings:
0, 1, 2, 3