The NIPC has a 4-pin connector for the room bus, a 5-pin connector for the speech module and a
5-pin connector for future use.
Each room bus offers individual addresses for up to four switch modules. The address is selected using
the DIP switch on the switch module printed circuit board.
The NIPC has a 2 meter long pull cord for call activation with two safety break plastic balls, and it is
available with three function buttons: red, green and yellow.
Important: At least one active NIPC module must be permanently located in the bathroom of each
resident apartment.
Note: The 4-pole connector terminal required for the room bus is not supplied with the
switch module. It is available as an accessory and must be ordered separately.
The NIPC requires a single backplate which must be ordered separately. The backplate enables this
switch module to be mounted over different types of back-boxes. Alternatively, a spacer with
installation kit is available for surface mounting the switch module.
6.8.1 Pull Cord Module (Active) Electrical Connections and DIP Switch Settings
This pull cord module is an “active peripheral” therefore it must be connected to one of the three
room buses of a teleCARE IP room controller by the 4-pole room bus connector (see 6.4.1, 4-Pole
Connector Terminal on page 65).
Each room bus offers individual addresses for up to four switch modules. The address is selected using
the DIP switch on the switch module printed circuit board.
For details of connecting a speech module refer to 6.15.1, Connecting the Speech Module on page
104
Figure 103. Pull cord module electrical connections and address setting
4-Pole Connector
for Room Bus
Dip Switch
for Setting the
Room Bus
Address
0, 1, 2, or 3
Connector
for the
Speech Module
Not supported
with UL 2560
systems