User's Manual
TD 93021US Rev A
10 FEB 2014 / Ver A 104
Installation Guide
teleCARE IP
• There must be a minimum distance of 1in (30mm) in all directions between the room
display and any metal objects to avoid the degradation of the card reader antenna.
6.15 Speech Module (NISP)
The Speech Module (NISP) is a teleCARE IP peripheral which facilitates speech in combination with
Ascom Interactive Messaging (IM). It consists of a gray or white plastic body, a printed circuit board
and a perforated anodized aluminum face plate.
Figure 135. Speech Module (NISP) front and back view
The NISP incorporates a pre-amp circuit and a speaker amplifier and includes two loudspeakers and an
electric microphone. It has a two-color LED which shows red to indicate that the speech direction from
staff to resident is active and green to indicate that the speech direction from resident to staff is active.
The NISP can only be used in combination with the teleCARE IP Door side Module (NIDM), the bedside
module (NIBM2), the Medical Rail Socket (NIMS2) the active Pull Cord Module (NICP), and the Room
Display (NIRD.
6.15.1 Connecting the Speech Module
The 5-wire cable with two 5-pole plugs is required to connect the speech module to the associated
switch module. It is available in three lengths: 7.0in (170mm) (R190192), 8.0in (200mm) (R190193)
and 15.75in (400mm) (660313). The 15.75in (400mm) long cable is used when the speech module is
mounted separately from the associated switch module.
For a two-module combination with the speech module mounted to the right side of the switch
module (viewed from the front) the R190192 cable 7.0in (170mm) is required.