User's Manual
TD 93021US Rev A
10 FEB 2014 / Ver A 82
Installation Guide
teleCARE IP
passive cancel switch module.
Figure 101. Door Side module electrical connections and address setting
The NIDM has a 5-pin connector for connecting a speech module. Refer to 6.15.1, Connecting the
Speech Module on page 104 for details of the speech module.
Note: The 4-pole connector terminals required for the room bus and the passive bus
connections module must be ordered separately.
6.8 Pull Cord Module - Active (NIPC-G3A and NIPC-W3A)
The Pull Cord Module (NIPC) is intended for use in the teleCARE IP system, in areas such as bathrooms
and toilets. It is an “active peripheral” therefore it must be connected to one of the three room buses
of a teleCARE IP room controller.
Figure 102. Pull cord module (active)
4-Pole
connector for
Passive Bus
4-Pole Connector
for Room Bus
Dip switch
for setting the
Room Bus
Address
0, 1, 2, or 3
Connector
for the
Speech Module
Buzzer