User's Manual

TD 93021US Rev A
10 FEB 2014 / Ver A 158
Installation Guide
teleCARE IP
Figure 178. Wiring for a 2-bed room with medical rail sockets
LAN
LAN
+24V
GND (0V)
+5V5
DATA
VOICE
GND
+5V5
DATA
VOICE
GND
+5V5
DATA
VOICE
GND
GND
GND
GND
+Vin
+Vout
+Vout
L1 L2
+5V5
DATA
VOICE
GND
1
+5V5
DATA
VOICE
GND
L1 L2
ON
1 2
0
1
2
3
1
ASCOM
+5V5
DATA
VOICE
GND
+5V5 o
IN0 o
IN1 o
OUT o
L1 L2
RELAYS
1
NIDM PASSIVE BUS
ON
1 2
0
1
2
3
1
ASCOM
+5V5
DATA
VOICE
GND
+5V5 o
IN0 o
IN1 o
OUT o
L1 L2
1
RELAYS
NIDM PASSIVE BUS
ON
1 2
0
1
2
3
1
ASCOM
+5V5
DATA
VOICE
GND
+5V5 o
IN0 o
IN1 o
OUT o
L1 L2
RELAYS
1
NIDM PASSIVE BUS
1
ASCOM
+5V5
DATA
VOICE
GND
+5V5 o
IN0 o
IN1 o
OUT o
L1 L2
1
RELAYS
NIDM PASSIVE BUS
L1
L2
L1
L2
Room Bus 0
Room Bus 1
Room Bus 2
LAN
Room Bus 1
Room Bus 0
Room Bus 1
Room Bus 0
Room Bus 0
Room Controller
NIRC
Doorside Module
NIDM
Pull Cord Module
NIPC (Active)
Pull Cord Module
NIPC (Active)
Medical Rail Socket
NIMS
Toilet Cancel Module
NITC (Active)
Light Switching
Circuits
Light Switching
Circuits
Medical Rail Socket
NIMS
Address 1
Address 4
(xed)
Address 0
Address 2
Address 2Address 1
Light Switching
Relay
Power Supply
Bi-stable
DC relay
To Bed
Light 1
Diode
- +
Light Switching
Relay
Power Supply
Bi-stable
DC relay
To Bed
Light 2
Diode
- +
L1 L2
0
1
2
3
ON
1 2
6.5mm
Max.
1mm²
1 2
+
Room Bus
Address Settings
Light Switching
Connector
Room Bus
Connector
Passive Bus
Connector
24Vdc
Connector
1
1
Cat.5 Cable
1
6.5mm
Max.
0.5mm²
1 2
Legend
+5.5V
IN0
IN1
OUT
+5.5V
DATA
VOICE
GND
1/4 in.
(
(
1/4 in.
(
(