Specifications
Signametrics 78
Return Value Integer error co
de.
Value
Meaning
ration successfully completed.
DMM_OKAY
Ope
Negativ
Example int status = DMMOpenCalACCaps(0);
DMMOpenTerminalCal
SMX2040 SM 2044
Description s.
Remarks This function characterizes the Inductance measurement path and source, which is
required prior to making inductance measurements. It should be performed within one
hour, before using the inductance measurements. For better accuracy it should be
performed more frequently. The Open Terminal calibration should be performed with the
test leads open. The DMMOpenTerminalCal sweeps the inductance stimulus source
across the full bandwidth, and makes measurements at several points. It takes about
twenty seconds to complete the process. For a complete characterization of the
Inductance measurement system it is also necessary to perform the inductance zero
operation with the inductance range and frequency selected, using the Relative function
and with the probes shorted.
Parameter
e Value
Error code
X2042 SMX ;
Calibrate the Inductance measurement function with open terminal
#include "sm204032.h"
int DMMOpenTerminalCal(int nDmm)
Type/Description
nDmm
int Identifies the DMM. DMMs are numbered starting with zero.
Return Value Integer error code.
Value
Meaning
DMM_OKAY
Operation successfully completed.
Negative Value
Error code
Example int status = DMMOpenterminalCal(0);
DMMPeriodStr
SMX2040 SMX2042 ; SMX2044 ;
Description Return the next DMM period reading, formatted for printing.
#include "sm204032.h"
int DMMPeriodStr(int nDmm, LPSTR lpszReading)