Specifications

119 Signametrics
Table 9-4 Resistance Test, 4-wire
Step Range Input Minimum Reading Maximum Reading
1
33 [1] 0 -2 m 2 m
1
33 [1] 10 9.9972 10.0028
1
330 0 -6 m 6 m
2
330 100 99.987 100.013
3
3.3 k 0 -33 m 33 m
4
3.3 k 1 k 0.999917 k 1.000083 k
5
33 k 0 -350 m 350 m
5
33 k 10 k 9.99905 k 10.00095 k
5
330 k 0 -5 5
6
330 k 100 k 99.986 k 100.014 k
[1] SMX2044 only.
Note: The use of 4-wire Ohms for resistance values above 300 k is not recommended.