Specifications
MODEL 370 REV: D Page 5 of 26
Equipment such as Scanning Monochromators
1.3 Accessories
The instrument is designed to accept all standard UDT Instruments Radiometric
/Photometric accessories including microphotometer and telephotometer
attachments, radiometric and photometric detector head assemblies, visible and
infrared LED measurement systems, integrating spheres for attenuation and
uniform response, detectors with smaller active areas, low profile head assemblies
for tight spaces, and special bandpass filters for UV through near infrared
measurements.
1.4 Calibration
Detector assemblies may be obtained with calibrations stored in a plug-in
calibration (PIC) EPROM. Stored information includes data for calibrations with
standard filters and accessories for radiometric or photometric measurements,
special filters, or specialized detector assemblies for measurements at particular
wavelengths such as 632.8nm for HeNe lasers, or continuous calibration data
covering a wide spectral range.
If accessories are obtained after initial purchase, either the PIC module can be
returned for calibration additions or a new one purchased.
Calibration data includes calibration certification listing model and serial numbers
of items calibrated, description of calibration including wavelength, measurement
units, special accessories or conditions, and a statement of traceability to NBS. UDT
Instruments maintains standards and equipment optimized for accurate calibration
and recalibration of instruments. The normal period to recalibration is six months.
An electrical calibration lab may have the facilities to check the current
measurement accuracy of the 370 (See performance verification). In this case only
the detector assemblies and PIC modules need be returned for calibration.
UDT Instruments keeps calibration data for each instrument on file.
2.0 PREPARATION FOR USE
2.1 Inspection
Before shipment this instrument was inspected and found to be free of defects.
When unpacking, inspect for damage that may have occurred during shipment. If
damage is found, file a claim with the carrier and notify the UDT Instruments
customer service department.
The Model 370 shipping container should contain:
1 – Model 370 Optometer
1 – Instruction Manual
1 – Carrying Case