Specifications

MODEL 370 REV: D Page 10 of 26
TABLE 1
IEEE CONTROL CODES
A Set Autorange
B0 restore screen
B1 blank screen
B2 Turn on backlight
B3 Turn off backlight
D set mode, DL log, DN linear, DR ratio, DG logratio.
E0 digital reading
E1 analog bar reading
F send current reading
Data is equal to input current divided by responsitivity
G Go
H Halt
K Display Caldata
Kx select Caldata x
M set Manual mode, does not change range
Mx set gain to specific exponent 3, 4, 5, 6, 7, 8, 9
R Send responsivity; R+x.xxxE+xx set responsivity
S send status of current channel to controller. Format is MODE WAVELENGTH
AUTO/MANUAL GAIN with each datapoint separated by space (ASCII 2Ōh)
U ask for Unit of measure
V1 -set unit to W
V2 - set unit to fc
V3 - set unit to Lux
V4 - set unit to f
V5 - set unit to W/cm2
V6 - set unit to W/cm2*sr
V7 - set unit to Cd/m2
V8 - set unit to Lm
WX set Wavelength 200 to 2000 depending on cal
WU Increments current wavelength by current step
WD Decrements current wavelength by current step
X display status of current channel
ZX set Zero mode, Z to turn on zero mode, Z1 to reset zero mode
*MESSAGE* displays “message” where message is an ASCII String of up to 16 characters