Specifications

76 Chapter4
Connectors
Rear Panel Connectors
Rear Panel Connectors
1. 10 MHz OUT
This port replaces the 10 MHz OUT port from the rear panel of the HP/Agilent 8922. The
signal is a general purpose 10 MHz reference output (sine wave).
10 MHz OUT is always active.
See Also
Specifications, Reference Specifications
2. 10 MHz IN
This port is always connected to the HP/Agilent 8922 rear panel 10 MHz OUT port, and
phase locks the down and up converter Local Oscillators in the Agilent 83220A/E to the 10
MHz reference from the HP/Agilent 8922. Connect together using the BNC cable provided
(Part No. 83220-61025).
NOTE
An independent 10 MHz reference should not be connected to the Agilent
83220A/E since the HP/Agilent 8922 can tune its 10 MHz reference for some
applications. The HP/Agilent 8922 10 MHz OUT and the Agilent 83220A/E 10
MHz IN should always be linked together.
5
1234