Specifications

Chapter 3 69
Screens
Tests
Tests
There is an additional selection added to the Procedure field. The new entry is SYS_CAL
and invokes a system calibration program.
SYS_CAL is a user-initiated calibration routine performed between the HP/Agilent 8922
and the Agilent 83220A. A system calibration should be initiated whenever a new
combination of HP/Agilent 8922 and Agilent 83220A are connected together, (or whenever
the HP/Agilent 8922 or Agilent 83220A are changed), or after a RAM Initialize from the
SERVICE screen.
NOTE
SYS_CAL can not be performed using the Agilent 83220E, due to the absence
of AUX ports.