Specifications

48 Chapter2
Verifying Performance
RF Analyzer Specifications
Output RF Spectrum Measurement
1
After one timeslot, 577 µs, from an isolated receiver hop trigger.
(When using “Output RF Spectrum due to Ramping” measurement, the dynamic range is
decreased by 12 dB (due to Peak Hold)).
Supplemental Characteristics
1. On an HP/Agilent 8922E/F/H, Output RF Spectrum Measurement is only available with Option
006.
Range
RF In/Out: 13 to +32 dBm.
Aux RF IN:(Agilent 83220A
only)
23 to +20 dBm.
Input Levels for Optimum
Dynamic Range
RF In/Out: 3, +2, +7, +12, +17, +22, +27, +32 dBm.
Aux RF IN:(Agilent 83220A
only)
13, 8, 3, +2, +7, +12, +17,+22 dBm.
Input Frequency Setting
Error
a
:
a. When setting up the RF analyzer for measurements, you enter the expected fre-
quency and level for the incoming signal. If the frequency is within the limits
shown the specified performance will result.
±10 kHz.
Input Level Setting Error: ±3 dB.
Log Linearity: ±0.4 dB.
Amplitude Flatness: ±1.0 dB.
Amplitude Resolution: 0.4 dB.
Dynamic Range (dB): This describes the spectrum analyzer resolution bandwidth filter
used when measuring output RF spectrum. The dynamic range
of the measurement will be a combination of this filter response
and the modulation spectrum of the incoming signal.
Offset (kHz)
100 200 300 400 600 800 to 1800
Range
(dB)
29 45 51 63 73 75
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com