User`s manual

4200-DR Series LaserSource User’s Manual · Page 7
Installation
Installation of the 4200-DR is very straightforward, as the quick start section
above illustrated. This section will provide additional details and considerations
for installing your 4200-DR.
After unpacking the unit, make sure all packing materials have been removed
and nothing obscures the ventilation ports on the side and bottom of the unit.
Changing the Voltage Selection
Before powering on the unit, ensure that the voltage selection on the IPC is set
correctly. Improper voltage selection can easily damage the unit. Changing the
voltage selection requires that you remove the voltage selection module from
the IPC. You can select from 100V, 120V, and 230V. Choose the voltage that
most closely matches your AC voltage.
Remove the power cord from the unit. Using a small, flat-blade screwdriver,
insert the tip of the screwdriver into the two small openings above and below the
voltage indication to release the voltage selection module from the IPC, as
shown in the picture below.
Input Power Connector (IPC)
Once the module has been removed, remove the small, white voltage selection
tumbler from the module, and then re-insert so that the desired voltage is
shown. Re-insert the module into the IPC.