4200-DR Series LaserSource User’s Manual · Page 31
Error Messages
Error
Code
Description Cause
E-100 General Error The error code is non-specific, and is generally
used when no other error code is suitable.
E-102 Message too long The message is too long to process.
E-123 Path not found The message used an invalid path command.
E-124 Data mismatch The message contained data that did not match
the expected format.
E-201 Data out of range The message attempted to set a value that was
outside the allowable range.
E-202 Invalid data type When trying to parse the message, the data was
in an invalid format.
E-204 Suffix not valid An invalid number base suffix (radix) was
encountered when parsing a number.
E-501 Interlock shutdown
output
The interlock input (pins 1 and 2 of the output
connector) were not shorted when the output was
on.
E-504 Laser current limit
disabled output.
The laser output was turned off because a current
limit was detected and the corresponding bit in
the OUTOFF register was set.
E-505 Laser voltage limit
disabled output
The laser voltage exceeded the voltage limit and
the output was turned off.
E-506 Laser photodiode
current limit disabled
output
The laser output was turned off because a
photodiode current limit was detected and the
corresponding bit in the OUTOFF register was
set.
E-507 Laser photodiode
power limit disabled
output
The laser output was turned off because a
photodiode power limit was detected and the
corresponding bit in the OUTOFF register was
set.
E-509 Laser short circuit
disabled output
The laser output was turned off because a short
condition was detected and the corresponding bit
in the OUTOFF register was set.
E-510 Laser out of tolerance
disabled output
The laser output was turned off because an out-
of-tolerance condition was detected and the
corresponding bit in the OUTOFF register was
set.
E-511 Laser control error
disabled output
A hardware control error was detected which
forced a shutdown of the laser output.
E-512 Power failure A power failure was detected.
E-514 Laser mode change
disabled output
A change in the operating mode of the
LaserSource while the output was on forced a
shutdown of the laser output.
E-516 Incorrect configuration
for calibration to start
The LaserSource was not configured properly,
including the mode and output on state, to be
able to start the desired calibration process.
E-517 Calibration must have
the output on to start
The laser output must be on for the calibration
process to start.