User`s manual
4200-DR Series LaserSource User’s Manual · Page 15
Menu
Description
Factory
Default
Advanced
Advanced Settings Menu
Off if Io Lim
Turns off the laser output if the Io limit is active No
Off if Im Lim
Turns off the laser output if the Im limit is exceeded Yes
Off if Po Lim
Turns off the laser output if the Po limit is
exceeded
Yes
Off if OutTol
Turns off the laser output if the laser goes out of
tolerance
No
User Cal
User Calibration Menu (located in Advanced Menu)
Edit Cal
Set to “Yes” to enable editing of any of the user
calibration entries
No
Reset Cal
Reset the TEC user calibration data to factory
defaults (all slopes to 1, all offsets to 0). Requires
Edit Cal to be set to Yes before this setting will be
displayed.
No
Various
M & B
Settings
Slope (M) and Offset (B) user calibration settings
for each set point and measurement. See the User
Cal - Field Calibration section below for additional
details on using slope/offset calibration.
No