User`s manual

Page 14 · 4200-DR Series LaserSource User’s Manual
Menu
Description
Factory
Default
On Delay
The delay, from the time the Output button is
pressed to when the output is actually energized.
3000ms
Comm Menu
Communications Menu
Set While
Rmt
To disable changing the set point while in Remote
Mode, set this value to “No”. To enable changing
the set point while in Remote mode, set this Value
to “Yes”.
No
Err While
Rmt
To turn off the display of errors while in remote
mode, set this value to “No”. To display errors
while in remote mode, set this value to “Yes”.
Yes
Terminal
Mode
Terminal mode simply echoes any characters
received over the USB interface.
No
Msg Term
This controls the output message termination, and
can be set to CR/LF, CR, LF, or None.
CR/LF
Sys Menu
System Settings Menu
Config
Allows for restoring the instrument to defaults,
saving configurations, and restoring
configurations. See Saving and Restoring
Configurations below for more information.
No Change
Display
Display mode, which is used to control the layout
on the main screen. Normal will display three
values, typically the set point current, photodiode
or power read back and voltage. Large will display
the read back current, depending on the mode (Io,
Im, Po, VF mode, respectively). When changing
the set point, the set point will be displayed during
the set point change.
Normal
Brightness
The vacuum florescent display can be set to one of
eight brightness levels.
100%
Audible
Beep
This setting controls when the unit produces
audible feedback. Set to No to prevent sound, or
Yes to enable audible alerts such an error
messages.
Yes
Lockout
Knob
Lockout knob allows you to disable knob
operation from the main display. This prevents
accidental changes of the set point. The knob will
always work in the menus regardless of this
setting.
No
Knob Speed
Controls the adjustment speed of the knob.
Possible values are slow, medium, and fast.
Slow