User`s manual

65
Current 1mA 1mA 1mA
Readback
Resolution
Voltage 1mV 2mV 4mV
Current 1mA 1mA 1mA
Meter
Resolution
Voltage 1mV 2mV 4mV
Current 1mA 1mA 1mA
Transient Response Time
Less than 1ms for output to recover to within 100 mV following a change in output
current from full load to half load or vice versa
Command Processing Time
Programming Commands: Maximum time for output to change after receiving APPLy
and SOURce commands: <50msec
Readback Command: Maximum time to readback output MEASure? command:
<100msec
The Other Commands: < 50msec
Supplemental Characteristics
Output Programming Range (maximum programmable values)
3662A 3663A 3664A
Voltage
035.2V 080.2V 0120.2V
Current
014.5A 06.5A 04.2A
Temperature Coefficient, ±(% of output + offset)
Maximum change in output/readback per °C after a 30-minute warm-up.
3662A 3663A 3664A
Voltage
30ppm + 0.5mV 30ppm+ 0.8mV 30ppm+ 1mV
Current
30ppm + 0.2mA 30ppm+ 0.1mA 30ppm + 0.1mA
Stability, ±(% of output + offset)
Following a 30-minute warm-up, the change occurs in output within 8 hours under
constant load, line, and ambient temperature.
3662A 3663A 3664A
Voltage
0.02% + 2mV 0.02% +3mV 0.02% + 4mV
Current
0.02% + 6mA 0.02% + 3mA 0.02% +2mA